Parts list – RKI Instruments 61-1001RK-05 User Manual
Page 15

61-1001RK-05 Combustible Gas Detector• 11
NOTE:
For convenience, leave regulator and calibration cup connected by the sample
tubing.
2.
When the controller display reading falls below the alarm points and then return the
controller to normal operation.
NOTE:
If you do not allow the gas reading to decrease below the alarm points, then
unwanted alarms may occur.
3.
Verify that the controller display reading decreases and stabilizes at 0 %LEL.
4.
Store the components of the calibration kit in a safe and convenient place.
Parts List
Table 3 lists replacement parts and accessories for the 61-1001RK-05 hydrogen detector.
Table 3: Parts List
Part Number
Description
18-0400RK-01
Junction box with spacers
NC-6205-05
Replacement hydrogen detector
71-0195RK
61-1001RK-05 Hydrogen Detector Operator’s Manual (this document)
81-0002RK-01
Calibration cylinder, 50 %LEL hydrogen in air, 34 liter steel
81-0076RK-01
Zero air calibration cylinder, 34 liter steel
81-1050RK
Regulator, 0.5 LPM with pressure gauge and flow control knob, for 17
and 34 liter steel calibration cylinders
81-1117RK
Calibration cup