Page 29

GX-94 Operator’s Manual

Updating the Gas Combination (2. COMBINE GAS)

Sets the display and circuitry for types of gases to be detected; there
are 11 possible combinations. Press AIR/

▲ or

(SHIFT)/

to scroll

through the combinations. Press ON/OFF/INPUT to accept the selected
combination.

Note

Change the GAS COMBINATION values only when the
appropriate sensors are installed in the corresponding sensor
sockets. Calibrate immediately after changing sensor combi-
nations.
Gases selected in this operation will be displayed throughout
all setup and normal display screens.

Viewing & Setting Alarms (3. SET ALARM)

Sets the alarm levels for gases detected by the GX-94. Press AIR/

to

change the alarm level for a single gas, or DISPLAY/(ADJ) to skip that
channel.

Press AIR/

to raise an alarm level, or (SHIFT)/

to lower the alarm

level. Hold down the button to scroll quickly through the alarm setting.
The lowest available setting for each alarm point is “off.”

Press ON/OFF/INPUT to accept the displayed alarm value and move to
the next alarm or gas.

SET ALARM

Allows the user to change LOW and HIGH HC alarms for
combustible gas.

SET ALARM

Allows the user to change the LOW and HIGH %O

2

alarms.

Remember that the LOW O

2

alarm is a decreasing alarm, for falling