8 • Operation

Model GX-86 Instruction Manual

Safety Self-Check

The GX-86 continuously monitors itself for proper operation. If a malfunction occurs, a
single steady “trouble” tone sounds, and one of the following messages display.

Emergency Operation

The GX-86 can operate temporarily with up to three failed sensors. If a sensor(s) fails or is
absent at start-up, the GX-86 sounds a seven-tone alert, displays

FAIL HC O2 HS CO

(any

combination of up to three) then adjusts to operate on the remaining sensor(s). The GX-86
displays

XX

for the defective sensor(s) and normal readings for the active sensors. The

GX-86 performs normal monitoring functions for the active sensors for 13 minutes, when
it performs a self-diagnosis, then shows a malfunction alarm. To continue using the
GX-86, press the OFF button then restart.

If a sensor fails during operation, the GX-86 shows a malfunction alarm. To continue
monitoring with the remaining active sensors, press the OFF button, and restart as
described above.

Table 4: GX-86 Error Messages

Display

Description

FAIL (CH4 O2 HS CO)

Indicates sensor failure of one or more sensors
(the failed sensor is displayed) before start up.

Element Trouble

Indicates combustibles sensor failure after
start up.

00 OVER 00.0

Indicates O

2

concentration in excess of 40.0% or

an O

2

sensor failure after start up.

WARNING: High oxygen concentration

increases the risk of fire hazard.

LOW BATT

Indicates the battery voltage is too low to run the
GX-86. See “Displaying battery voltage” on
page 6.