Maintenance – RKI Instruments GP-01 User Manual
Page 26

21 • Maintenance
Model GP-01 Operator’s Manual
Maintenance
This section describes troubleshooting procedures for the Model GP-01. It
also describes how to change or charge the Model GP-01’s batteries as
well as how to replace the unit’s gas sensor and filter disks.
WARNING: RKI Instruments, Inc. recommends that service,
calibration, and repair of RKI instruments be performed
by personnel properly trained for this work. Replacing
sensors and other parts with original equipment does not
affect the intrinsic safety of the instrument.
Troubleshooting
The troubleshooting table describes error messages, symptoms, probable
causes, and recommended actions for problems you may encounter with
the Model GP-01.
Table 6: Troubleshooting the Model GP-01
Symptoms
Probable
Causes
Recommended Action
• The LCD is
blank.
• The unit may
have been
turned off.
• The alkaline
batteries may
need to be
replaced or the
Ni-cad
batteries
recharged.
1. To turn on the unit, press
and hold the POWER/
MODE button.
2. If the unit does not turn on,
replace the alkaline
batteries or recharge the
Ni-cad batteries.
3. If the difficulties continue,
contact RKI Instruments,
Inc. for further instruction.
• The LCD shows
abnormally high
readings but
other gas
detection
instruments do
not.
• The unit may
need to be
recalibrated.
• The sensor
may need
replacement.
1. Recalibrate the unit.
2. Replace the sensor and
calibrate the unit.
3. If the difficulties continue,
contact RKI Instruments for
further instruction.