Restarting the as400 – Retsch AS 400 User Manual
Page 15

06.11.2012
Retsch GmbH
15
Operating the AS400 control
Switching on and off
E
The main switch E is located at the left-front under the AS400
control.
Turn the main switch on
On illuminates in the memory display
050 illuminates in the speed/“g“ display
The interval off LED illuminates
Two bars illuminate in the time display
The AS400 control is now ready for operation, without intermit-
tent mode, in continuous duty, at a speed of 50 r.p.m..
Automatic recognition of imbalance
4
5
15
16
The AS400 is equipped with a function which automatically recognises any
imbalance.
Even running of the AS400 is monitored by imbalance sensors. In case of an
acceleration >=0.3g, the sensors switch off the drive unit of the AS400.
Imbalance triggering characteristic
Display 5 flashes
The green LED above button 15 extinguishes
The red LED above button 16 lights up
The AS400 can no longer be started.
Reset – restarting the AS400
Display 5 flashes
Press button 4 to reduce the speed until display 5 no longer flashes
The AS400 can now be started again by pressing button 15 once more.
Necessary reductions :
201 – 300 reduce to 200
151 – 200 reduce to 150
101 – 150 reduce to 100
76 – 100 reduce to 75
51 – 75 reduce by 1 rpm
Speeds of <050 are not possible