Retsch SR 300 User Manual
Page 9

2/8/2012
Retsch GmbH
9
0002
8
9
1
2
3
4
5
6
7
Fig. 2
Installation site – Altitude
max. 2000 m above mean sea level
Setting up the unit
Set up the SR 300 on a solid surface.
In order to bolt the base frame to the floor (maximum 10 mm screw
diameter), it will be necessary to remove the two adjusting screws 6
and the plastic caps 7 at the front. Fig. 2
Electrical connection
The power connection may be made only by a qualified and licensed
electrician.
The voltage and frequency specifications for the SR 300 will be
found on the machine’s data plate.
Ensure that the values shown there correspond to those for the
local power supply.
Use the supplied power cord to connect the SR 300 to the power
network.
The line power cord supplied does not have a plug since the design
of the plug will depend on the installation location and the regula-
tions applicable in the particular country.
When connecting the power cord to the line supply, external fus-
ing shall be provided in accordance with local codes.
Failure to observe the values on the data plate can cause damage to
either the electrical or the mechanical components or both.
It will be necessary to check the direction of rotation (corresponding
to the arrow on the motor housing) before putting the machine into
service for the first time.
If the rotation direction is reversed, grinding will not be satisfactory
and mechanical components could be damaged.