9 fault messages, Fault messages – Retsch SM 300 User Manual
Page 37
Fault messages
37
9 Fault messages
F.01
Overload
Frequency converter switched off because of overloading
F.04
Open the door
Close the door and close the lock.
F.04.- Open the lock
Close the lock
F.11
Engine is overheated.
Starting impossible. Please wait until it cools down.
F.15
Fault in the safety circuit of the
frequency converter activation
Door switch opened,
Handle switch opened,
The motor-driven lid lock monitoring has detected a fault.
F.18
Fault in th
e transformer’s safety
circuit
Contact monitoring Relay Inrush current limiter
F.26
Excessive temperature
frequency converter
Frequency converter is overheated
F.37
No communication with the
frequency converter
Frequency converter does not respond
(4 seconds after start)
See also other documents in the category Retsch Equipment:
- BB 50 (41 pages)
- BB 100 (24 pages)
- BB 200 (24 pages)
- BB 300 (24 pages)
- Cryomill (68 pages)
- CryoMill Autofill 50l (28 pages)
- DM 200 (20 pages)
- DM 400 (60 pages)
- Emax (67 pages)
- GM 200 (44 pages)
- GM 300 (44 pages)
- MM 200 (32 pages)
- MM 400 (32 pages)
- PM 100 (44 pages)
- PM 400 (36 pages)
- PM GrindControl (48 pages)
- RM 200 (36 pages)
- RS 200 (32 pages)
- SK 100 (24 pages)
- SM 100 (48 pages)
- SM 200 (48 pages)
- Cyclone - SM 300 (20 pages)
- SR 200 (24 pages)
- SR 300 (24 pages)
- TWISTER (44 pages)
- ZM 200 (32 pages)
- AS 200 jet - Automatic Vacuum Regulation (6 pages)
- AS 200 control - natural frequency (32 pages)
- AS 200 (36 pages)
- AS 200 jet (56 pages)
- AS 200 jet (48 pages)
- AS 200 tap 30.025.0001 (28 pages)
- AS 200 tap 30.025.1001 (28 pages)
- AS 300 (32 pages)
- AS 400 (32 pages)
- AS 400 (28 pages)
- AS 450 basic (40 pages)
- AS 450 basic (44 pages)
- AS 450 control (52 pages)
- DR 100 (40 pages)
- DR 100 (28 pages)
- PP 25 (32 pages)
- PP 40 (28 pages)
- PT 100 (36 pages)