Operating the machine
21
•
Close the waste gas / overflow valve (V3).
Pos : 6.18 /00005 Ü berschriften/1.1 Ü bersc hriften/1.1 Übersc hriften BD A/11 Druc kaufbau @ 1\mod_1240400926229_9.doc @ 8240 @ @ 1
4.7 Pressure build-up
Pos : 6.19 /00010 Bedi enungsanleitung en Kapitels ammlungen/Cr yoMill/0016 LN 2 D ewar 50l/1604 Modul Dewar 50l Druc kaufbau @ 2\mod_1267701477789_9.doc @ 18841 @ @ 1
•
Close the withdrawal valve (V1) and the waste gas / overflow valve (V3).
•
Open the pressure build-up valve (V2).
•
Watch the pressure indicator (DM).
•
Close the pressure build-up valve (V2) as soon as the needle reaches the
red marking or nitrogen comes out of the pressure relief valve (PSV).
Fig. 6: Pressure gauge
Pos : 6.20 /00005 Ü berschriften/1.1 Ü bersc hriften/1.1 Übersc hriften BD A/11 Entnahme von Flüs sigstic kstoff @ 2\mod_1267774957314_9.doc @ 18877 @ @ 1
4.8 Withdrawal of liquid nitrogen
Pos : 6.21 /00010 Bedi enungsanleitung en Kapitels ammlungen/Cr yoMill/0016 LN 2 D ewar 50l/1605 Modul Dewar 50l Entnahme Flüssigstic kstoff @ 2\mod_1267709072517_9.doc @ 18855 @ @ 1
•
Attach the liquid nitrogen line to the CryoMill and to the connection (SA).
•
Open the filling and withdrawal valve (V1).
•
After taking the liquid, close the filling and withdrawal valve (V1) again.
Pos : 6.22 /00005 Ü berschriften/1.1 Ü bersc hriften/1.1 Übersc hriften BD A/11 Druc kentl astung @ 2\mod_1267777113490_9.doc @ 18891 @ @ 1
4.9 Pressure relief
Pos : 6.23 /00010 Bedi enungsanleitung en Kapitels ammlungen/Cr yoMill/0016 LN 2 D ewar 50l/1606 Modul Dewar 50l Druc kentlastung @ 2\mod_1267709470983_9.doc @ 18862 @ @ 1
•
Close the pressure build-up valve (V2).
•
Open the waste gas valve (V3).
Pos : 6.24 /00005 Ü berschriften/1.1 Ü bersc hriften/1.1 Übersc hriften BD A/11 Trans port @ 0\mod_1226495164391_9.doc @ 3398 @ 2 @ 1
4.10 Transport
Pos : 6.25 /00004 Warnhinweis e/V0019 VOR SICHT Überdr uc k i m Behälter - Gefärdung durc h LN 2 @ 2\mod_1267775099915_9.doc @ 18884 @ @ 1
CAUTION
Risks of liquid or gaseous nitrogen
Positive pressure in the vessel
–
The introduction of energy during transport can cause a sudden surge in
pressure. Nitrogen is blown off then and this can cause a lack of oxygen,
cryogenic burns, oxygen enrichment and associated risks.
•
Make sure the vessel is depressurised before you transport it,
•
Remove the siphon and put in the transport plugs.
•
Transport the vessel only in an upright position.
•
Avoid impacts and vibrations.
•
Secure the vessel against falling down, rolling away and damage.
Pos : 6.26 /00005 Ü berschriften/1. Ü berschriften/1 Prüfanweis ungen @ 1 \mod_1245230896441_9.doc @ 11114 @ @ 1
Pos : 6.27 /00010 Bedi enungsanleitung en Kapitels ammlungen/Cr yoMill/0016 LN 2 D ewar 50l/1612 Modul Dewar 50l Verboten ist die An wendung folgender Mittel und @ 2\mod_1268231540473_9.doc @ 18957 @ @ 1
It is impermissible to use the following agents and tools:
•
pliers
•
striking tools
DM