4.3 Technical Data
Pos : 6.11 /00010 Bedi enungsanleitung en Kapitels ammlungen/Cr yoMill/0016 LN 2 D ewar 50l/1611 Modul Dewar 50l Tec hnische D aten @ 2\mod_1268230808121_9.doc @ 18950 @ @ 1
Type: APOLLO® 50
Overall height: 80cm
Outside diameter: 50cm
Overall width: 76cm
Immersion depth 58.5cm
Neck diameter: 5.0cm
Weight empty: 44kg
Weight full: 85 kg
Geometrical capacity: 49.5 litres
Stat. rate of evaporation: 2% / day
Operating pressure: max. 1.3 bar
Pos : 6.12 /00005 Ü berschriften/1.1 Ü bersc hriften/1.1 Übersc hriften BD A/11 Vakuumversc hlus s und Sicherheits einric htung @ 2\mod_1268297292227_9.doc @ 18970 @ @ 1
4.4 Vacuum closure and safety device
Pos : 6.13 /00010 Bedi enungsanleitung en Kapitels ammlungen/Cr yoMill/0016 LN 2 D ewar 50l/1613 Modul Dewar 50l Vakuumversc hluss und Sicherheits einric htung @ 2\mod_1268297689386_9.doc @ 18977 @ @ 1
Fig. 3: Vacuum closure and safety device
WARNING
The vacuum closure and the safety device protect the vacuum space from positive
pressure . Only the manufacturer’s qualified staff may conduct the re-evacuation .
CAUTION
The protective cap (2) retains the valve insert (1) if excessive pressure builds up in
the vacuum space.
• Do not remove the protective cap (2).
• Protect the valve from heat or cooling because embrittlement can lead to the loss
of operating vacuum.
Pos : 6.14 /00005 Ü berschriften/1.1 Ü bersc hriften/1.1 Übersc hriften BD A/11 Fülls tandsanz eige @ 2\mod_1268300595137_9.doc @ 18991 @ @ 1
4.5 Level Display
Pos : 6.15 /00010 Bedi enungsanleitung en Kapitels ammlungen/Cr yoMill/0016 LN 2 D ewar 50l/1614 Modul Dewar 50l Füllstands kur ve Apollo ® 50 @ 2\mod_1268298441800_9.doc @ 18984 @ @ 1