Temperature accuracy, Warranty, Model 522 operating instructions – Practical Instrument Electronics 522 Rev D User Manual
Page 4

Model 522 Operating Instructions
82 E. Main Street Suite 3.14 · Webster, NY 14580
Tel: 585-872-9350 · Fax: 585-872-2638 [email protected] · http://www.piecal.com/
Practical Instrument Electronics, Inc. Copyright ã 2011. All rights reserved.
522-9002 Rev D
4-4
Open T/C Test Pulse
< 1 uA for 300 ms
Open T/C Response Time
< 3 seconds
Open T/C Threshold
10 kW nominal
Available Options:
Carrying Case
Part Number: 020-0205
T/C Wire Kits
Part Number: 020-0202 wire kit 1 includes J, T, E, K, leads with mini connectors
Part Number: 020-0203 wire kit 2 includes B, R/S, N leads with mini connectors
Temperature Accuracy
The following charts give worst-case temperature accuracy based on stated millivolt accuracy of ±(0.008 % + 0.006). Temperature
is uncompensated on the horizontal axis, referenced to 0 °C. Cold Junction calibration accuracy of 0.1 °C is not included in the
temperature error.
TYPE B
594 to 1820 °C
0
0.2
0.4
0.6
0.8
1
1.2
500
700
900
1100
1300
1500
1700
1900
Temperature °C
Er
ro
r °
C
TYPE E
-260 to 1000 °C
0
0.2
0.4
0.6
0.8
1
1.2
1.4
-400
-200
0
200
400
600
800
1000
1200
Temperature °C
Er
ro
r °
C
TYPE J
-210 to 1200 °C
0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
0.4
-400
-200
0
200
400
600
800
1000
1200
1400
Temperature °C
Er
ro
r °
C
TYPE K
-245 to 1372 °C
0
0.2
0.4
0.6
0.8
1
1.2
-400
-200
0
200
400
600
800
1000
1200
1400
1600
Temperature °C
Er
ro
r °
C
TYPE N
-229 to 1300 °C
0
0.2
0.4
0.6
0.8
1
1.2
-400
-200
0
200
400
600
800
1000
1200
1400
Temperature °C
Er
ro
r °
C
TYPE R
24 to 1768 °C
0
0.2
0.4
0.6
0.8
1
1.2
0
200
400
600
800
1000
1200
1400
1600
1800
2000
Temperature °C
Er
ro
r °
C
TYPE S
21 to 1768 °C
0
0.2
0.4
0.6
0.8
1
1.2
0
200
400
600
800
1000
1200
1400
1600
1800
2000
Temperature °C
Er
ro
r °
C
TYPE T
-251 to 400 °C
0
0.2
0.4
0.6
0.8
1
1.2
-300
-200
-100
0
100
200
300
400
500
Temperature °C
Er
ro
r °
C
Warranty
Our equipment is guaranteed against defective material and workmanship (excluding batteries) for a period of three years from the
date of shipment. Claims under guarantee can be made by returning the equipment prepaid to our factory. The equipment will be
repaired, replaced or adjusted at our option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our
guarantee. No responsibility is accepted for damage, loss or other expense incurred through sale or use of our equipment. Under no
condition shall Practical Instrument Electronics, Inc. be liable for any special, incidental or consequential damage.