Practical instrument electronics, Accessories, Thermocouple ranges & accuracies – Practical Instrument Electronics 422 User Manual
Page 8
![background image](/manuals/570442/8/background.png)
Page 8
Practical Instrument Electronics
82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Accessories
Optional: T/C Wire Kit 1 for Types J, K, T & E
Part No. 020-0202
Optional: T/C Wire Kit 2 for Types B, R/S & N
Part No. 020-0203
Three feet (1 meter) of T/C extension wire, stripped on one end with a miniature T/C
male connector on the other end.
Standard Test Leads (Included with calibrator)
Part No. 020-0207
Three feet (1 meter) of wire with an alligator clip on one end and a banana plug on the other end.
Optional Ni-MH 1 Hour Charger w/4 Ni-MH AA Batteries
Part No. 020-0103
(100-120 V AC input for North America Only)
General
Accuracy
±(0.008% of Reading + 0.006 mV)
Cold Junction Compensation
± 0.09°F (±0.05 °C) - Thermistor traceable to NIST for 11 years
Millivolt Range
-13.000 to 80.000 mV
Operating Temperature Range
-25 to 60 °C (-10 to 140 °F)
Temperature Effect
≤ ±50 ppm/°C; Cold Junction Sensor ≤ ±25 ppm/°C
Relative Humidity Range
10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing
Size
L=5.63 x W=3.00 x H=1.60 inches
Weight
12.1 ounces (including boot & batteries)
Batteries
Four “AA” Alkaline 1.5V (LR6)
Battery Life
50 Hours
Optional NiMh Rechargeable battery kit
120 VAC for North America Only; charger, four NiMh
batteries, AC & DC cords [Part # 020-0103]
Low Battery
Low battery indication with nominal 1 hour of life left
Protection against misconnection
Over-voltage protection to 60 V dc (rated for 30 seconds)
Display
High contrast graphic liquid crystal display. LED backlighting
for use in low lit areas.
Read
Input Impedance
> 10 Megohms
Open Thermocouple
Threshold
Pulse
10,000 Ohms nominal
< 10 microamp pulse for 300 milliseconds
Normal Mode Rejection
50/60 Hz, 50 dB
Common Mode Rejection
50/60 Hz, 120 dB
Source
Output Impedance
< 0.3 Ohms
Source Current
> 20 mA (drives 80 mV into 10 Ohms)
Noise
≤ 4 microvolts p-p for frequencies of 10 Hz or below
Thermocouple Ranges & Accuracies
Unless otherwise indicated all specifications (except Cold Junction) are rated from a nominal 23 °C, 70 % RH for 1 year from calibration