Model 512 operating instructions, Temperature accuracy, Warranty – Practical Instrument Electronics 512 User Manual
Page 4

Model 512 Operating Instructions
841 Holt Road #1 • Webster, NY 14580
Tel: 585-872-9350 • Fax: 585-872-2638
[email protected] • http://www.piecal.com/
Practical Instrument Electronics, Inc. Copyright 2003. All rights reserved.
512-9002 2/25/04
4-4
Temperature Accuracy
The following charts give worst-case temperature accuracy based on stated resistance accuracy of ±(0.015 % + 0.05) Ω.
Pt100 TCR=0.00385
-200 to 850 °C
0.1
0.2
0.3
0.4
-200
0
200
400
600
800
Temperature °C
Error °C
Pt100 TCR=0.003902
-200 to 850 °C
0.1
0.2
0.3
0.4
-200
0
200
400
600
800
Temperature °C
Error °C
Pt100 TCR=0.003916
-200 to 850 °C
0.1
0.2
0.3
0.4
-200
0
200
400
600
800
Temperature °C
Error °C
Pt100 TCR=0.003926
-200 to 850 °C
0.1
0.2
0.3
0.4
-200
0
200
400
600
800
Temperature °C
Error °C
Cu10 TCR=0.00427
-200 to 260 °C
1.2
1.25
1.3
1.35
1.4
-200
-100
0
100
200
300
Temperature °C
Error °C
Ni110 TCR=0.00530
-100 to 300 °C
0.08
0.1
0.12
0.14
0.16
-100
0
100
200
300
400
Temperature °C
Error °C
Ni120 TCR=0.00672
-80 to 260 °C
0.07
0.08
0.09
0.1
-100
-50
0
50
100
150
200
250
300
Temperature °C
Error °C
Warranty
Our equipment is guaranteed against defective material and workmanship (excluding batteries) for a period of three years from the
date of shipment. Claims under guarantee can be made by returning the equipment prepaid to our factory. The equipment will be
repaired, replaced or adjusted at our option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our
guarantee. No responsibility is accepted for damage, loss or other expense incurred through sale or use of our equipment. Under no
condition shall Practical Instrument Electronics, Inc. be liable for any special, incidental or consequential damage.