Practical Instrument Electronics

82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com

Page 3

Configuration

Configure the Calibrator

Move

w

POWER SWITCH to

“SOURCE” or

“READ”.

Setup
Double click the

e

DIAL KNOB at any time the

unit is on and the following displays will appear
for 15 seconds:

Turn the

e

DIAL KNOB to move through the

menu. Press the

e

DIAL KNOB to toggle

between OFF and ON or to scroll through the
settings.

MODEL 51#B V#.##

DOUBLE CLICK

EZ-DIAL KNOB

FOR CONFIGURATION

> EXIT

15

TEMP UNITS °C

T/C

K

AUTO OFF

ON

EXIT MENU - exits this menu immediately and
saves any changes. Menu will automatically exit
after 15 seconds of inactivity (countdown timer is
displayed).
TEMP UNITS - pressing the knob will toggle
between °C and °F.

T/C -
520: pressing the knob will toggle between the
factory configured T/C Type and mV.

521: pressing the knob will cycle through T/C
types J, T, E, K, R, S, B, N, G, C, D, L (J-DIN), U
(T-DIN), Platinel II and mV.

AUTO OFF - If AUTO OFF is ON, the unit will
turn off after 30 minutes of inactivity to save
battery life. If AUTO OFF is OFF the unit will stay
on until the POWER SWITCH is moved to the off
position.

Note: All settings are remembered even with the
power off. Removing the batteries resets the
values to factory defaults.

Simulating thermocouples requires the use of
thermocouple or extension grade thermocouple
wire.

Plug thermocouple wires into the female
miniature thermocouple connector mounted in
the top end of the housing.

The PIECAL 520B/521B has two banana jacks
mounted in the top end of the housing. These
are not temperature compensated and are to be
used only for millivolt signals.

Connections

Push & Hold

to

Store/Ramp

Push & Turn

for

Fast Dialing

Double Click

for Options

HI

TYPE K °C

Thermocouple Wire

Device

Requiring

Calibration

OFF

THERMOCOUPLE SOURCE

SOURCE

HI

SET

LO

J‐T‐E‐K‐R‐S‐B‐N‐G‐C‐D‐L‐U‐P‐mV

TYPES

322.1

(Optional Boot Shown)

Banana jacks and female thermocouple

connector on the top of the calibrator