beautypg.com

Page 20, 2, 3 & 4 wire rtd connections – Practical Instrument Electronics 820 User Manual

Page 22

background image

Page 20

RTD Source

Choose this function to provide a simulated RTD signal into
controllers, temperature transmitters, indicators or any input
devices that measure RTD sensors.

Move the power switch

w

to SOURCE then Double Click the

EZ-DIAL knob to get into the Double Click Menu. Turn the knob
to scroll through the settings and press the knob to make your
selection. Select RTD for the FUNCTION, °F or °C for the UNITS
and RTD (Choose from one of Platinum 100Ω, or 1000Ω, Copper
10Ω or 50Ω, Nickel 120Ω or 110Ω curves). Note: Pt 100Ω 3850
is the most common RTD type.

Disconnect all sensor wires from the devices to be calibrated
and connect the PIECAL 820 to the inputs of the device using 2,
3 or 4 wires.

Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO. You may also select
any third output setting (such as mid-range) using the SET
position on the EZ-CHECK switch. The output is adjusted in 0.1°
increments by turning the knob

e

. Press and turn the knob for

faster dialing with 10.0° increments.

2, 3 & 4 Wire RTD Connections

Process Calibrator

mA ∙ V ∙ TC ∙ Ω ∙ RTD ∙ Hz

SOURCE

HI

SET

LO

OFF

MAX
READ

MIN

READ

OVERLOAD

Hz SYNC

Double Click

Menu

Push & Turn

for

Fast Dialing
Push & Hold

to

Store/Step

READ

Pt 100

α

=3850 °C

820.0

HI

Instrument with RTD Input

Controller

Temperature Transmitter

Temperature Indicator

Temperature Trip or Alarm