Page 6
RTD
Type
Alpha
Degrees C
Range
Accuracy
°C
Degrees F
Range
Accuracy
°F
Pt 100 Ohm (DIN/IEC/JIS 1989)
Based on ITS-90
1.3850
(0.00385)
-200.0 to 200.0
200.0 to 600.0
600.0 to 850.0
±0.3°
±0.4°
±0.5°
-328.0 to 392.0
392.0 to 1112.0
1112.0 to 1562.0
±0.5°
±0.7°
±0.8°
Pt 100 Ohm (Burns)
1.3902
(0.003902)
-195.6 to 200.0
200.0 to 648.9
±0.3°
±0.4°
-320.0 to 392.0
392.0 to 1200
±0.5°
±0.7°
Pt 100 Ohm (Old JIS 1981)
1.3916
(0.003916)
-200.0 to 200.0
200.0 to 648.9
±0.3°
±0.4°
-328.0 to 392.0
392.0 to 1200
±0.5°
±0.7°
Pt 100 Ohm (US Lab)
1.3926
(0.003926)
-200.0 to 100.0
100.0 to 700.0
700.0 to 862.2
±0.3°
±0.4°
±0.5°
-328.0 to 212.0
212.0 to 1292.0
1292.0 to 1584.0
±0.5°
±0.7°
±0.8°
Pt 1000 Ohm
1.3850
(0.00385)
-200.0 to 200.0
200.0 to 266.0
±0.3°
±0.4°
-328.0 to 392.0
392.0 to 511.0
±0.5°
±0.7°
Pt 1000 Ohm Hy-Cal HVAC
1.3750
(0.00375)
-200.0 to 200.0
200.0 to 274.0
±0.3°
±0.4°
-328.0 to 392.0
392.0 to 525.0
±0.5°
±0.7°
Copper 10 Ohm (Minco)
1.4274
(0.004274)
-200.0 to 260.0
±2.1°
-328.0 to 500
±4.1°
Copper 50 Ohm
1.4280
(0.00428)
-150.0 to 50.0
±1.1°
-58.0 to 302.0
±1.1°
Ni 120 Ohm (Pure)
1.6720
(0.00672)
-80.0 to -315.6
±0.2°
-112.0 to 600.0
±0.3°
Ni 110 (Bristol 7 NA)
1.5801
(0.005801)
-106.7 to 315.6
±0.3°
-160.0 to 600.0
±0.3°
Ranges & Accuracies