Common Specifications for all Current Modes:

Ranges

0.000 to 24.000 mA, -25.00 to 125.00% of 4-20 mA
0.000 to 52.000 mA, -25.00 to 105.00% of 10-50 mA

Accuracy

≤ ± (0.025 % of reading + 0.004 mA)

Temperature Effect

≤ ± 50 ppm/°C of range

Resolution

0.001 mA and 0.01 %

Step/Ramp Timebase Accuracy

0.01% of 4.9152 MHz


Source/Power and Measure 2-Wire Transmitter Specifications:

Loop Compliance Voltage

≥ 43 Volts

Loop Drive Capability

1200 Ω at 20 mA/800 Ω at 50 mA for entire battery life

Miscellaneous

Open loop or out of compliance conditions are indicated by appropriate error display
Battery life in:
Source mode ≥ 18 hrs at 12 mA/≥ 9 hrs at 30 mA typical
Power measure ≥ 10 hrs at 12 mA/≥ 5 hrs at 30 mA typical
Selectable EZ-Step(s) for Source Mode/2-Wire Transmitter Simulation:
2 to 16 selectable step settings
Step size is determined by the selected high & low ranges
Selectable time settings for stepping and soak:
STEP: 5 to 900 seconds SOAK: 0 to 900 seconds


Read mA Specifications:

Voltage Burden

≤ 2V

Overload/Current Limit Protection

≤ 24 mA (4-20 range)/ ≤ 52.5 mA (10-50 range) nominal

Battery Life

≥ 40 hours typical


2-Wire Transmitter Simulation Specifications:

Overload/Current Limit Protection ≤ 24 mA (4-20 range)/ ≤ 52.5 mA (10-50 range) nominal
Loop Voltage Limits

2-90 VDC

Miscellaneous

Open loop or out of compliance conditions are indicated by appropriate error display
Battery life ≥ 40 hour typical
Selectable EZ-Step(s) for Source Mode/2-Wire Transmitter Simulation:

2 to 16 selectable step settings

Step size is determined by the selected high & low ranges
Selectable time settings for stepping and soak:

STEP: 5 to 900 seconds SOAK: 0 to 900 seconds