H.
Manual Step and Auto Step/Ramp Parameter
To Change the EZ-Step
ä Size:
1.
Press and hold the SOURCE/STEP SIZE button for more than
¾ of a second.
2.
The display will flash “EZ-STEP SIZE”.
3.
Turn the EZ-Dialä knob to select from 2 to 16 steps between
the EZ-Checkä limits.
4.
Turn the EZ-Dialä clockwise past 16 steps to select continuous
ramp mode.
5.
Press the SOURCE/STEP SIZE button again to return to the
normal display.
Note: If the EZ-Step
ä
option is turned off, the display will flash “EZ-STEP
OFF”. Refer to Model 532 Configuration, section B.
To Change the EZ-Step
ä Time:
1.
Press and hold the UNITS/STEP TIME button for more than
¾ of a second.
2.
The display will flash “EZ-STEP TIME”.
3.
Turn the EZ-Dialä knob to select from 5 to 900 second ramp
time. The time per step is calculated based on the selected EZ-Stepä
size.
4.
Press the SOURCE/STEP SIZE button to return to the normal
display.
To Change the Soak Time:
1.
Press and hold the READ/SOAK TIME button for more than
¾ of a second.
2.
The display will flash “SOAK TIME”.
3.
Turn the EZ-Dialä knob to select from 0 to 900 second soak
time.
Note: A soak time of 0 defeats the soak period. The step time will be
used instead.
4.
Press the READ/SOAK TIME button again to return to the
normal display.
To Change the EZ-Step
TM
Direction:
1. Press and release the EZ-Dialä knob without turning.
2. The display will change to show the EZ-Stepä direction selected (
or
).