Door alignment – MARVEL Scientific 25CM User Manual
Page 6

DOOR ALIGNMENT
Figure 5
Hinge adapter bracket lo-
cated on the top and bottom
of the door.
Remove top
hinge pin to re-
move the door.
Door must be
parallel to top
and sides of
refrigerator.
9/32”
(7.1mm)
NOTE
6
For the door to close properly, it is necessary to maintain a
minimum space of 9/32” (7mm) between the door and cabinet
flange (See Figure 5). This space can be adjusted by adjusting the
top and bottom hinge adapter brackets.
Door Alignment Procedure
The door should be parallel to the sides and top of the refrig-
erator. If alignment is necessary the door may be adjusted by
loosening the 2 screws which secure the hinge adapter brackets
on the top and bottom of the door and adjusting the door side
to side. Use a 5/32” allen wrench for this procedure. (See Figure
5 below). When finished aligning the door, tighten the screws
securely.