Labnet Z400K High Capacity Centrifuge User Manual
Page 20

20 E
Z 400 K / March 1997
Error no.: 11
Error no.: 11
Error no.: 11
Error no.: 11
Error no.: 11
Temperature sensor
Temperature sensor
Temperature sensor
Temperature sensor
Temperature sensor
Reason:
Short circuit at the temperature sensor or at the sensor cable
Action:
Call service
Reason:
Chamber temperature is too low, below - 25°C
Action:
Solenoid valve is not working. Call service
Error no.: 20
Error no.: 20
Error no.: 20
Error no.: 20
Error no.: 20
No rotor identification
No rotor identification
No rotor identification
No rotor identification
No rotor identification
Reason:
No rotor inserted
Action:
Insert rotor into the unit
Reason:
Rotor identification sensor defective
Action:
Call service
Reason:
Inserted rotor has no indicator ring
Action:
Use a correct rotor
Reason:
Rotor is not fixed correctly to the motor shaft
Action:
Insert rotor correctly. The pins have to align correctly with the
rotor slots (see chapter 3)
Error no.: 21
Error no.: 21
Error no.: 21
Error no.: 21
Error no.: 21
Start Bit is missing
Start Bit is missing
Start Bit is missing
Start Bit is missing
Start Bit is missing
Reason:
A magnet of the indicator ring is missing
Action:
Check indicator ring and call service
Error no.:
Error no.:
Error no.:
Error no.:
Error no.: 10
10
10
10
10
Overtemperature in the rotor chamber (more than + 50 ° C)
Overtemperature in the rotor chamber (more than + 50 ° C)
Overtemperature in the rotor chamber (more than + 50 ° C)
Overtemperature in the rotor chamber (more than + 50 ° C)
Overtemperature in the rotor chamber (more than + 50 ° C)
Reason:
Breakdown of the refrigeration system / Electronic failure
Action:
Call service
Reason:
Temperature sensor defective / Electronic failure
Action:
Call service