Labnet Z400K High Capacity Centrifuge User Manual
Page 14

14 E
Z 400 K / March 1997
4.7
Preselection of brake intensity and acceleration
With key "fast/slow" (1) you have the choice between two steps of brake and
acceleration intensity.
1
2 3
Figure 8
If the green control lamp "fast" (2) is lighting, the unit accelerates and decelerates
fast.
If the green control lamp "slow" (3) is lighting, the unit accelerates and decelerates
very softly.
Acceleration and deceleration time of all rotors
rotor
deceleration time
acceleration time
in sec.
in sec.
fast
slow
fast
slow
220.27 V02
35
69
43
50
220.59 V06
25
45
16
35
220.81 V05
38
88
40
71
220.86 V01
62
125
66
86
220.88 V01
33
57
23
37
220.89 V01
23
37
18
32
220,92 V01
30
57
21
37
220.93 V01
62
135
58
88
220.96 V02
34
90
28
63
220.97 V02
33
89
28
62
221.02 V01
34
60
34
60
221.03 V01
33
63
28
56
fast
slow
preset
speed