Abnormality & troubleshooting, Abnormality reason action – Kinetik Blood Pressure BPM1C User Manual
Page 16
Abnormality & Troubleshooting
Abnormality
Reason
Action
The cuff does not inflate or the display
shows low pressure during inflation
LCD screen shows battery symbol
LCD screen shows abnormal result
LCD shows “EE”
No response when buttons pressed
Pressure loss due to air leakage in
tube or cuff
Rubber tube is not connected
properly
Batteries are low
Cuff is not tight or is
positioned incorrectly
Subject was excited/nervous or
spoke/moved during measurement
Wrong position
Some interference during inflation
or incorrect operation during
measurement
Insufficient inflation
Some interference during inflation
or incorrect operation during
measurement
Incorrect operation
Check the connection between the tube
and monitor
Reconnect the rubber tube properly
Change the batteries - refer to the ‘Loading
the Batteries’ section
Tighten the cuff and refer to the section
‘Setting up your Blood Pressure Monitor’
Re-test. Do not speak and take 2-3 deep
breaths before to relax. Stay calm and
remain steady.
Adjust your position - refer to part a on
page 7
Refer to the section ‘Using your Blood
Pressure Monitor’
Re-measure
Re-measure
Take out the batteries for a few minutes,
then re-insert them, following directions in
section ‘Loading the Batteries’.
15