KDS Microliter OEM Module
12
HYPER-TERMINAL: Terminal Emulation Software (continued)
You may want to save the setup information under a descriptive filename.
4. At the PC, launch HyperTerminal with the above setup specifications (if it is not
already running). Type PROM? at the PC keyboard and verify that the pump
module's version is displayed at the PC terminal.
5. Type RUN to start the pump; type STOP to stop the pump. After starting the
pump, > should be displayed, indicating pump is infusing. After stopping the
pump, : should be displayed.