6

KDS EZFlow 2010/2011

Operation Instructions

1)

CHECK THE “DOSE CALIBRATION” SWITCH LOCATED ON THE BOTTOM OF

THE PUMP EACH TIME YOU USE A NEW/DIFFERENT SIZE/BRAND SYRINGE.

“Dose Calibration” switch is located on the bottom of the pump. This switch can

be used to offset variances between syringes from different manufacturers as

follows:

IMPORTANT: ONLY ONE SWITCH CAN BE SET IN THE ‘UP’ POSITION. IF TWO OR

MORE SWITCHES ARE SET IN THE ‘UP’ POSITION, THE PUMP WILL REVERT BACK

TO THE ‘STD’ SETTING

2)

Pump can be used either on any horizontal surface or on the IV pole with pole

clamp.

3)

Plug power cord into a wall outlet or use the DC 12V built-in battery power. (If

using battery power, charge battery at least 10 hours before use. A fully charged

battery can continuously support running up to 4 hours at a flow rate of 50 ml/hr.)

4)

Turn the power switch on. ON/OFF switch is located at the right side of the pump.

When initialized, the pump will go through a Self Diagnostic program. After the

pump passes the Self Diagnostic tests, “0.1” will appear on the display. If ‘ERR’

shows on the display the pump did not pass the power-on test and should not be

used until the malfunction is corrected.

5)

Press “Number Set Key” (up & down arrows) to set desired flow rate.

6)

Load syringe with desired fluid. Lift up the Syringe Securing Bar

and rotate right 90˚. Place the syringe into the bed. Be sure to insert the nozzle

into the syringe slot. Then press the release button on the plunger driver block and

slide the driver block into position so that the end of the syringe plunger can be

placed into the slot in the driver block. Make sure the syringe plunger is properly

seated in the plunger driver block. Turn the Syringe Securing Bar to the left 90

degrees to secure the syringe barrel. Make sure the syringe barrel is secure. Press

and hold the PURGE button to purge air from the syringe and associated tubing.

7)

Press “RUN” to start infusion.

8)

To reset the total volume delivered, power off the pump and restart.

BD

Monoject

Terumo

20ml

30ml

50/60ml

20ml

50/60ml

20ml

50/60ml

+20%

STD

+5%

+5%

+20%

+10%

STD