Keypad Function:

(Continued)

KEYPAD DESCRIPTION

AUDIBLE AND VISUAL ALARMS

INCREASE rate - Increases flow rate.

Press Hold to change the rate faster.

DECREASE rate - Decreases flow rate.

Press Hold to change the rate faster.

CHECK VOLUME - Press and hold RUN mode to display the volume delivered.

Release to display the preset flow rate.

PURGE - Only functions in STOP mode. Press to run the pump at high speed

expelling air out of the syringe or to purge the syringe. Rapid purging rate varies

according to syringe size:

Syringe size (ml)

20

50/60

100

Max. flow rate (ml/hr)

150

300

300

RUN - Press RUN button to start infusion; the LED indicator lights flash.

STOP - Press to stop pump

MUTE - Press to silence the alarm buzzer while `NEAR EMPTY’, `COMPLETION’ or

`OCCLUSION’ alarm is sounded. Audible and visual alarms.

NEAR EMPTY - When there is approximately 5 ml of medication left, an audible

and visual alarm is activated. Press the MUTE button to silence the alarm buzzer.

COMPLETE - When infusion is complete, the operation stops and audible and visual

alarms are activated. Press MUTE button to silence the alarm buzzer.

LOW BATT - Audible and visual alarms are activated if the battery runs low on

charge. To clear the larm, plug the power lead into and appropriate AC wall outlet.

the pump will continue to operate while recharging the battery.

BATTERY POWER indicator light (orange) is on to indicate that built-in bettery power

is in use.

AC POWER indicator light )green) is to indicate the pump is connected to main AC

power and is charging battery.

OCCLUSION - Whilst occlusion occurs, operation stops. Press MUTE button to

silence the alram buzzer. Infusion only continues after the alarm condition is cleared.

KDS EZFlow 2010/2011