Operation instructions, Storage – KD Scientific EZFLOW 2020 User Manual

Page 7

background image

6

KDS EZFlow 2020/2021

Operation Instructions

1)

Load the syringe with the desired fluid. Lift up the syringe securing knob and

rotate towards the right 90˚. Place the syringe into the bed. Insert the nozzle into

the syringe slot. Press and hold the secure button on the driver to bring the driver

block into a position that the extremity of the syringe plunger can be placed into

the driver slot. Turn the securing knob to the left 90˚, and secure the syringe.

2)

The ON/OFF switch at the left side of the pump is a power switch. Connect the

power cable to a wall outlet and turn the pump on. Press and hold the “PURGE”

button to purge the syringe and the set tubing.

3)

Press the “SET TIME” button to preset the desired infusion time, while the green

light indicates a corresponding number is on.

4)

Press the “RUN/STOP” button to start infusion; the running green light will be on.

During the infusion, the “SET TIME” and “RAPID” buttons are locked. If you need

to adjust the infusion time setting, press “RUN/STOP” followed by “SET TIME”

button to reset the infusion time. To continue operation, press “RUN/STOP” again.

5)

During the infusion, when “RUN/STOP” is pressed, the pump stops operation

and goes into a HOLD mode with operation indicator (green) on. To continue

the rest of infusion with the original setting, simply press “RUN/STOP” again.

NOTE:

a)

If there is an adjustment on the syringe or a resetting of the infusion time,

the pump will begin infusion with the new settings. Otherwise, when

“RUN/STOP” is pressed, the pump will continue the rest of infusion with

the original settings.

b)

If the syringe or the tubing becomes occluded beyond the pump’s

capability to pump, the alarm will activate and the OCCLUSION

indicator (red) will come on. The pump will stop operating.

c)

To remove the syringe, lift up the syringe securing knob and rotate

right 90˚. The RUN indicator goes off when the syringe is removed.

Storage

This product should be stored in an environment with ambient temperature range of

14˚F-131˚F, (-10˚C-55˚C) relative humidity below 93% and air pressure

500hpa~1060hpa. Do not store close to corrosive vapor and/or harmful impurities.