Keypad function – KD Scientific EZFLOW 2020 User Manual
Page 6
5
KDS EZFlow 2020/2021
Keypad Function:
(Continued)
KEYPAD DESCRIPTION
ALARMS AND INDICATOR LIGHTS
SET TIME - Activated only in STOP mode. When pressed, sets the desired time. `SET
TIME’ indicator should illuminate.
PURGE - Only functions in STOP mode. When pressed, starts purging tubing set.
RUN/STOP - Press to start or stop the pump. `MODE’ indicator should illuminate.
NEAR END - When there is approximately one minute of infusion time remaining,
this indicator flashes and the alarm buzzer sounds. To silence the alarm, press the
`RUN/STOP’ button.
COMPLETE/OCCLUSION - When infusion is complete, the indicator is illuminated in
red and the alarm buzzer sounds.
When the device senses an occlusion by not being able to drive with a force of
6.5kg ± 1 kg, the indicator is illuminated in red and the alarm buzzer sounds. To
silence the alarm, press the `RUN/STOP’ button.
The battery indicator light (ornage) indicates the driver is using built-in battery power.
this light and AC power light (green) will blink together if battery is low.
AC power light (green) indicates the pump is connected to mains AC power and is
charging the battery.