KDS EZFlow 2020/2021

Keypad Function:

(Continued)

KEYPAD DESCRIPTION

ALARMS AND INDICATOR LIGHTS

SET TIME - Activated only in STOP mode. When pressed, sets the desired time. `SET

TIME’ indicator should illuminate.

PURGE - Only functions in STOP mode. When pressed, starts purging tubing set.

RUN/STOP - Press to start or stop the pump. `MODE’ indicator should illuminate.

NEAR END - When there is approximately one minute of infusion time remaining,

this indicator flashes and the alarm buzzer sounds. To silence the alarm, press the

`RUN/STOP’ button.

COMPLETE/OCCLUSION - When infusion is complete, the indicator is illuminated in

red and the alarm buzzer sounds.

When the device senses an occlusion by not being able to drive with a force of

6.5kg ± 1 kg, the indicator is illuminated in red and the alarm buzzer sounds. To

silence the alarm, press the `RUN/STOP’ button.

The battery indicator light (ornage) indicates the driver is using built-in battery power.

this light and AC power light (green) will blink together if battery is low.

AC power light (green) indicates the pump is connected to mains AC power and is

charging the battery.