beautypg.com

Technical specifications – JULABO SL-14K Calibration Baths User Manual

Page 9

background image

SL

9

2.4. Technical

specifications

Calibration Baths

SL-8K

SL-14K

Working temperature range

°C

50 ... 300

50 ... 300

Temperature stability

(1

°C

±0.005

±0.005

Temperature uniformity

°C <±0.008

<±0.008

Overall dimensions (WxDxH)

cm 21x42x46

21x42x60

Bath opening (WxL)

cm

12

12

Bath depth

cm 17

31

Filling volume

Liter

14

24

Weight

kg 48

51




SL C

Temperature selection

digital

via

keypad

indication on LCD DIALOG-DISPLAY (°C/°F)

remote control via PC

indication on monitor

Temperature indication

VFD COMFORT-DISPLAY (°C/°F)

Resolution

°C

0.01

ATC3

INT / EXT

°C

±3 / ±9

Temperature control

ICC - Intelligent Cascade Control

Heater wattage (at 230 V)

kW

3.0

Heater wattage (at 115 V)

Electronically adj. pump capacity

stages

1 ... 4

Flow rate

at 0 bar

l/min

22 ... 26

Max. pressure

at 0 liters

bar

0.7

Max. suction

at 0 liters

bar

0.4

Ambient temperature

°C

5 ... 40

Mains power connection 230 V/50-60 Hz

V/ Hz

190 ... 253 / 50-60

Current draw

(at 208 V / 230 V)

A

13 / 14

Mains power connection 115 V/50-60 Hz

V/ Hz

-----

Current draw

(at 100 V / 115 V)

A

-----




All measurements have been carried out at: rated voltage and frequency

ambient temperature: 20 °C

bath fluid: water

operating temperature: 70 °C

Technical changes without prior notification reserved.