Counter-cooling – JULABO SL-14K Calibration Baths User Manual

Page 21

background image

SL

21

5.4. Counter-cooling

Caution:

Securely attach all tubing to prevent slipping.

Observe the laws and regulations of the water distribution company valid in the
location where the unit is operated.

T16A

T16A

ext

Pt100

SERIAL

ALARM

REG+
E-PROG

STAND-BY

20

For applications near the ambient temperature, the cooling coil (20)
must be connected to the water mains.

 Using tubing, connect the cooling coil to the tap water supply,

and lead the tap water in a sink through the return connector.

A specific water flow rate of 750 ml/minute (15 °C) is sufficient
to compensate for the characteristic temperature.

For applications near ambient temperature (20 °C), the cooling
water temperature should at least be 5 °C below the working
temperature.

Controlled countercooling
Use the MVS solenoid valve controller
(26) and a solenoid valve (27) for
controlled countercooling to reduce tap
water consumption.

The circulator recognizes the MVS
controller via the socket

(10) and

automatically adjusts the necesssary
parameters.

Accessory:

Order No.

Description

9 790 000

MVS controller for

cooling

water

8 980 700

Solenoid valve