Code 14, Code 15, Code 16 – JULABO LH 85 High Dynamic Temperature System User Manual
Page 60: Code 17, Code 31, Code 03, Code 04, Code 20, Code 21, Code 40
Troubleshooting guide / Error messages
60
CODE 14
Safety sensor defect >TANK<
The safety temperature value lies below the working temperature
setpoint. Set the safety temperature to a higher value.
The temperature system is operated without bath liquid, or the liquid
level is insufficient. Replenish the bath tank with the bath liquid.
Tube breakage has occured (insufficient filling level due to excessive
bath liquid pumped out). Replace the tubing and replenish the bath tank
with the bath liquid.
The float is defect (e. g., because damaged in transit). Repair by
authorized JULABO service personnel.
CODE 15
External control selected, but external Pt100 sensor not connected.
CODE 16
Heating circuit interrupted.
CODE 17
Heating circuit short-circuited.
CODE 31
Communication between control module RD and instrument interrupted.
Warning
CODE 40
Warnings without a complete shutdown of the unit
CODE 03
Excess temperature warning
CODE 04
Low temperature warning
CODE 20
Cooling of the condenser is affected. Clean air-cooled condenser.
Check the flow rate and cooling water temperature on water-cooled
condenser.
CODE 21
Compressor stage 1 does not work.
Check mains circuit breakers F1 and F3, both might have triggered.
Reset circuit breakers if necessary (see below).
CODE 40
Low liquid level warning in the internal reservoir
CODE 41
High liquid level warning in the internal reservoir. (LH85)
These messages appear every 10 seconds as long as the compressor
is not switched on although requested by the instrument.
After a short cooling interval, the motor will be automatically
reconnected and the message
"
CODE XX
"
no longer appears.