JULABO FP51-SL Ultra-Low Refrigerated Circulators User Manual
Page 84
Remote control
84
Command
Parameter
Response of circulator
out_mode_03
1
Set external programmer input to current.
Current
0 mA ... 20 mA
out_mode_04
0
Temperature control of internal bath.
out_mode_04
1
External control with Pt100 sensor.
out_mode_05
0
Stop the unit = R –OFF-.
out_mode_05
1
Start the unit.
out_mode_08
0
Set the control dynamics - aperiodic
out_mode_08
1
Set the control dynamics - standard
out_sp_00
xxx.xx
Set working temperature. “Setpoint 1”
out_sp_01
xxx.xx
Set working temperature. “Setpoint 2”
out_sp_02
xxx.xx
Set working temperature. “Setpoint 3”
out_sp_03
xxx.xx
Set high temperature warning limit “OverTemp”
out_sp_04
xxx.xx
Set low temperature warning limit “SubTemp”
out_sp_06
xxx.xx
Set manipulated variable for heater via serial interface
-100 ... +100 [%]
out_sp_07
x
Set the pump pressure stage. (1 ... 4)
out_par_04
x.x
CoSpeed for external control 0 ... 5.0.
out_par_06
xxx
Xp control parameter of the internal controller.
0.1 ... 99.9
out_par_07
xxx
Tn control parameter of the internal controller.
3 ... 9999
out_par_08
xxx
Tv control parameter of the internal controller.
0 ... 999
out_par_09
xxx
Xp control parameter of the cascade controller.
0.1 ... 99.9
out_par_10
xxx
Proportional share of the cascade controller.
1 ... 99.9
out_par_11
xxx
Tn control parameter of the cascade controller.
3 ... 9999
out_par_12
xxx
Tv control parameter of the cascade controller.
0 ... 999
out_par_13
xxx
Maximum internal temperature of the cascade controller.
out_par_14
xxx
Minimum internal temperature of the cascade controller.
out_par_15
xxx
Band limit (upper)
0 ... 200 °C
out_par_16
xxx
Band limit (lower)
0 ... 200 °C
out_hil_00
-xxx
Set the desired maximum cooling power (0% to 100%).
Note: Enter the value with a preceding negative sign!
This setting is meaningful only for FP cooling machines.
out_hil_01
xxx
Set the desired maximum heating power (10% to 100%).