JULABO FP51-SL Ultra-Low Refrigerated Circulators User Manual
Page 22
Preparations
22
Backflow safety device
T16A
T16A
T1.25A
ext
Pt100
SERIAL
ALARM
REG+
E-PROG
STAND-BY
13
14
(see page 66
If the liquid levels in the circulator bath and the external
system are at different heights, overflowing must be
prevented after the power has been turned off.
Flood hazard!
For this reason, solenoid valves for loop circuit or shut-off
valves can be integrated in the loop circuit.
Connect the valve to the connector (13).
If socket (14) is used:
In menu >Inputs/Outputs< set the
menu item >SV-Out< to >Backflow<.
Order No. Description
8 980 701 Set of solenoid valves
for loop circuit up to +100 °C
8 970 456 Shut-off valve (suitable up to +90 °C)
8 970 457 Shut-off valve (suitable up to +200 °C)
5.3.1. Tubing
Recommended tubing:
Order No.
Length
Temperature range
8930008 1
m
CR
®
tubing
8 mm inner dia.
-20 °C to 120 °C
8930012 1
m
CR
®
tubing
12 mm inner dia.
-20 °C to 120 °C
8930108
1 m
Viton tubing
8 mm inner dia.
-50 °C to 200 °C
8930112
1 m
Viton tubing
12 mm inner dia.
-50 °C to 200 °C
8930410
1 m
Insulation for tubing 8 mm inner dia.
-50 °C to 100 °C
8930412
1 m
Insulation for tubing 12 mm inner dia.
-50 °C to 100 °C
8 930 209
0.5 m
8 930 210
1.0 m
8 930 211
1.5 m
8 930 214
3.0 m
Metal tubing, triple insulated, M16x1
-100 °C to +350 °C
8 930 220
0.5 m
8 930 221
1.0 m
8 930 222
1.5 m
8 930 223
3.0 m
Metal tubing, insulated, M16x1
-50 °C to +200 °C
Warning:
Tubing:
At high working temperatures the tubing used for temperature application and cooling
water supply represents a danger source.
A damaged tubing line may cause hot bath fluid to be pumped out within a short time.
This may result in:
Burning of skin
Difficulties in breathing due to hot atmosphere