11.3. List of commands

out commands: Setting parameters or temperature values.

Command

Parameter

Response of recirculating cooler

out_mode_05

0

Stop the unit = R –OFF-.

out_mode_05

1

Start the unit.

out_sp_00 xxx.xx

Set working temperature

in commands: Asking for parameters or temperature values to be displayed.

Command

Parameter

Response of recirculating cooler

version

none

Number of software version (V X.xx)

status none

Status message, error message (see page 29 )

in_pv_00

none

Actual bath temperature.

in_sp_00 none

Working

temperature

in_mode_05 none

Recirculating cooler in Stop/Start condition:
0 = Stop
1 = Start

11.4. Status

messages

Status messages

Description

00 MANUAL STOP

Recirculating cooler in „OFF“ state.

01 MANUAL START

Recirculating cooler in keypad control mode.

02 REMOTE STOP

Recirculating cooler in „r OFF“ state.

03 REMOTE START

Recirculating cooler in remote control mode.

11.5. Error

messages

Error messages

Description

-01 LOW LEVEL ALARM

Low liquid level alarm.

-05 WORKING SENSOR ALARM

Working temperature sensor short-circuited or
interrupted.

-03 EXCESS TEMPERATURE WARNING

High temperature warning. Starting at 75 °C
(no deactivation)
The return temperature soon reaches the switch-off
value of the high temperature warning function (85 °C)

-07 I

2

C-BUS ERROR

Internal error when reading or writing the I

2

C bus.

-08 INVALID COMMAND

Invalid command.

-09 COMMAND NOT ALLOWED IN

CURRENT OPERATING MODE

Invalid command in current operating mode.

-10 VALUE TOO SMALL

Entered value too small.

-11 VALUE TOO LARGE

Entered value too large.