beautypg.com

JULABO FLW20006 Recirculating Coolers User Manual

Page 30

background image

30

11.2. Draining

Notice:

Store and dispose the used bath fluid according to the laws for environmental
protection.

Risk of injury for hands when mounting the venting grid.

The unit has two drain taps. Both have to be used.
First drain the large tank via drain tap A, the drain the
remaining liquid via drain tap B.


 Turn off the unit and disconnect the mains cable from

the power source.

 Hold the venting grid, pull out and remove.
 Place a suitable vessel for accepting the used bath

liquid underneath the drain.

 Assemble the barbed fitting to the drain tap.

Slide a short piece of tube onto the barbed fitting and
hold it into a pail.

 Open the drain tap and empty the unit completely.

Tighten the drain tap after draining the unit and

replace the venting grid.

This manual is related to the following products: