JULABO FLW20006 Recirculating Coolers User Manual
Page 14

14
Return flow safety device
If the filling volume of the bath tank is not sufficient, prevent the liquid from flowing
back by using shut-off valves..
Order No.
Description
Suitable for
8 970 458
Shut-off valve G 1¼“
FL(W) 7006/11006/20006
The following questions shall help to recognize possible dangers and to reduce the
risks to a minimum.
Are all tubes and electrical cables connected and installed?
Note:
sharp edges, hot surfaces in operation, moving machine parts, etc.
What to do when a dangerous substance was spilled on or in the unit?
Before starting to work, obtain information concerning the substance and determine the
method of decontamination.
4.1. Tubing
Caution:
Employ suitable connecting tubing.
Make sure that the tubing is securely attached.
Avoid sharp bends in the tubing, and maintain a sufficient distance from
surrounding walls.
Regularly check the tubing for material defects (e.g. for cracks).
Preventive maintenance: Replace the tubing from time to time.
Recommended tubing:
Order No.
Description
Suitable for
8930325
1 m Reinforced tubing 1“ inner dia. (-40 ... +120°C)
FL(W) 7006/11006/20006
Tubing insulation
8930425
1 m Insulation, 35 mm inner dia.
Reinforced tubing 1“ inner dia.
Tube clamps
8970484
2 Tube clamps, size 5
Reinforced tubing 1“ inner dia.