background image

Operator responsibility – Safety recommendations

16


FP40-HE

FP50-HE
FPW50-HE

Working temperature range

°C

-40 ... 200

-50 ... 200

Temperature stability

°C

±0.01

±0.01

Cooling capacity

Medium: ethanol

°C
kW

+20 0 -20 -30
0.68 0.5 0.32 0.04

+20 0 -20 -40
0.9 0.8 0.5 0.16

Refrigerant

R134a

R404A or R507

Overall dimensions

(WxDxH) cm

36x46x71

42x49x72

Bath opening

(WxL) cm

23x14

18x12

Bath depth

cm

20

20

Filling volume

liters

9 ... 16

5.5 ... 8

Weight

kg

49

57

Mains power connection

230 V/50 Hz

V/Hz

207-253 / 50

207-253 / 50

Current draw

(at 230 V)

A

13

14

Mains power connection

230 V/60 Hz

V/ Hz

207-253 / 60

207-253 / 60

Current draw

(at 230 V)

A

13

14



HE

Temperature selection

digital

via keypad

indication on VFD COMFORT-DISPLAY

remote control via PC

indication on monitor

Temperature indication

VFD COMFORT-DISPLAY

Resolution

°C

0.01

ATC3

INT/EXT

°C

±3 / ±9

Temperature control

ICC - Intelligent Cascade Control

Heater wattage (at 230 V)

kW

2,0

Heater wattage (at 115 V)

kW

1,0

Electronically adj. pump capacity

stages

1 ... 4

Flow rate

at 0 bar

l/min

22 ... 26

Max. pressure at 0 liters

bar

0.7

Max. suction at 0 liters

bar

0.4

Electrical connections:

See page 17

Ambient temperature

°C

5 … 40





All measurements have been carried out at:
rated voltage and frequency

ambient temperature: 20 °C

Technical changes without prior notification reserved.