JULABO FPW50-HE Refrigerated and Heating Circulators User Manual
Page 16

Operator responsibility – Safety recommendations
16
FP40-HE
FP50-HE
FPW50-HE
Working temperature range
°C
-40 ... 200
-50 ... 200
Temperature stability
°C
±0.01
±0.01
Cooling capacity
Medium: ethanol
°C
kW
+20 0 -20 -30
0.68 0.5 0.32 0.04
+20 0 -20 -40
0.9 0.8 0.5 0.16
Refrigerant
R134a
R404A or R507
Overall dimensions
(WxDxH) cm
36x46x71
42x49x72
Bath opening
(WxL) cm
23x14
18x12
Bath depth
cm
20
20
Filling volume
liters
9 ... 16
5.5 ... 8
Weight
kg
49
57
Mains power connection
230 V/50 Hz
V/Hz
207-253 / 50
207-253 / 50
Current draw
(at 230 V)
A
13
14
Mains power connection
230 V/60 Hz
V/ Hz
207-253 / 60
207-253 / 60
Current draw
(at 230 V)
A
13
14
HE
Temperature selection
digital
via keypad
indication on VFD COMFORT-DISPLAY
remote control via PC
indication on monitor
Temperature indication
VFD COMFORT-DISPLAY
Resolution
°C
0.01
ATC3
INT/EXT
°C
±3 / ±9
Temperature control
ICC - Intelligent Cascade Control
Heater wattage (at 230 V)
kW
2,0
Heater wattage (at 115 V)
kW
1,0
Electronically adj. pump capacity
stages
1 ... 4
Flow rate
at 0 bar
l/min
22 ... 26
Max. pressure at 0 liters
bar
0.7
Max. suction at 0 liters
bar
0.4
Electrical connections:
See page 17
Ambient temperature
°C
5 … 40
All measurements have been carried out at:
rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.