beautypg.com

JULABO F38-EH Refrigerated and Heating Circulators User Manual

Page 13

background image

EH

13

Notice:

Check the safety installations at least twice a year!

 Excess

temperature

protection

according to IEC 61010-2-010.

With a screwdriver turn back the adjustable excess temperature protection
until the shut-down point (actual temperature).

 Low level protection according to IEC 61010-2-010.

To check the function of the float, it can be manually lowered with a
screwdriver for example.

Caution:

The temperature controlling i.e. of fluids in a reactor constitutes normal circulator
practice.

We do not know which substances are contained within these vessels.
Many substances are:
 inflammable, easily ignited or explosive
 hazardous

to

health

 environmentally

unsafe

i.e.: dangerous

The user alone is responsible for the handling of these substances!
The following questions shall help to recognize possible dangers and to reduce
the risks to a minimum.
 Are all tubes and electrical cables connected and installed?

Note:
sharp edges, hot surfaces in operation, moving machine parts, etc.

 Do dangerous steams or gases arise when heating?

Is an exhaust needed when working?

 What to do when a dangerous substance was spilled on or in the unit?

Before starting to work, obtain information concerning the substance and
determine the method of decontamination.