JULABO F34-EH Refrigerated and Heating Circulators User Manual
Page 9
EH
9
2.4. Technical
specifications
F12-EH
F25-EH
Working temperature range
°C
-20 ... 150
-28 ... 150
Temperature stability
°C
±0.03
±0.03
Temperature selection
digital
digital
Temperature indication
LED
LED
Resolution °C
0.1
0.1
ATC - Absolute Temperature Calibration 1-point 1-point
Temperature control
PID1
PID1
Heater wattage (at 230 V)
kW
2.0
2.0
Heater wattage (at 115 V)
kW
1.0
1.0
Cooling capacity
°C
+20 0 -20
+20 0 -20
Medium ethanol
kW
0.16 0.1 0.02
0.26 0.2 0.06
Refrigerant
R134a
R134a
Circulating pump:
discharge, max.at 0 bar
l/min
15
15
pressure, max. at 0 l
bar
0,35
0,35
Electrical connections:
External alarm device
Vdc/mA
24-0 / max. 25
24-0 / max. 25
Overall dimensions (WxDxH)
cm
20x36x56
23x42x61
Bath opening (WxL)
cm
13x15
12x14
Bath depth
cm
13
14
Filling volume from ... to
liters
3 ... 4.5
3 ... 4.5
Weight kg
22
30
Ambient temperature
°C
5 ... 40
5 ... 40
Mains power connection 230 V/50 Hz
V/ Hz
207-253 / 50
207-253 / 50
Current input (at 230 V)
A
11
12
Current input (at 230 V) CH
A
<9 + 1>
<9 + 2>
Mains power connection 230 V/60 Hz
V/ Hz
207-253 / 60
207-253 / 60
Current input (at 230 V)
A
11
12
Mains power connection 115 V/60 Hz
V/ Hz
103-127 / 60
103-127 / 60
Current input (at 115 V)
A
12
13
Mains power connection 100 V/50-60 Hz
V/ Hz
90-110V/50-60Hz
90-110V/50-60Hz
Current input (at 100 V)
A
15
13
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.