beautypg.com

JULABO F34-EH Refrigerated and Heating Circulators User Manual

Page 11

background image

EH

11



F33-EH

F34-EH

Working temperature range

°C

-30 ... 150

-30 ... 150

Temperature stability

°C

±0.03

±0.03

Temperature selection

digital

digital

Temperature indication

LED

LED

Resolution °C

0.1

0.1

ATC - Absolute Temperature Calibration 1-point

1-point

Temperature control

PID1

PID1

Heater wattage (at 230 V)

kW

2.0

2.0

Heater wattage (at 115 V)

kW

1.0

1.0

Cooling capacity

°C

+20 0 -20

+20 0 -20

Medium ethanol

kW

0.5 0.32 0.12

0.45 0.32 0.14

Refrigerant

R134a

R134a

Circulating pump:

discharge, max.at 0 bar

l/min

15

15

pressure, max. at 0 l

bar

0.35

0.35

Electrical connections:

External alarm device

Vdc/mA

24-0 / max. 25

24-0 / max. 25

Overall dimensions (WxDxH)

cm

36x46x69

38x58x62

Bath opening (WxL)

cm

23x14

24x30

Bath depth

cm

20

15

Filling volume from ... to

liters

12 ... 16

14 ... 20

Weight kg

43

41

Ambient temperature

°C

5 ... 40

5 ... 40

Mains power connection 230 V/50 Hz

V/ Hz

207-253 / 50

207-253 / 50

Current input (at 230 V)

A

12

12

Mains power connection 230 V/60 Hz

V/ Hz

207-253 / 60

207-253 / 60

Current input (at 230 V)

A

12

13

Mains power connection 115 V/60 Hz

V/ Hz

103-127 / 60

103-127 / 60

Current input (at 115 V)

A

15

14

Mains power connection 100 V/50-60 Hz V/

Hz 90-110V/50-60Hz ------

Current input (at 100 V)

A

15

------


All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C

Technical changes without prior notification reserved.