JULABO F34-EH Refrigerated and Heating Circulators User Manual
Page 11
EH
11
F33-EH
F34-EH
Working temperature range
°C
-30 ... 150
-30 ... 150
Temperature stability
°C
±0.03
±0.03
Temperature selection
digital
digital
Temperature indication
LED
LED
Resolution °C
0.1
0.1
ATC - Absolute Temperature Calibration 1-point
1-point
Temperature control
PID1
PID1
Heater wattage (at 230 V)
kW
2.0
2.0
Heater wattage (at 115 V)
kW
1.0
1.0
Cooling capacity
°C
+20 0 -20
+20 0 -20
Medium ethanol
kW
0.5 0.32 0.12
0.45 0.32 0.14
Refrigerant
R134a
R134a
Circulating pump:
discharge, max.at 0 bar
l/min
15
15
pressure, max. at 0 l
bar
0.35
0.35
Electrical connections:
External alarm device
Vdc/mA
24-0 / max. 25
24-0 / max. 25
Overall dimensions (WxDxH)
cm
36x46x69
38x58x62
Bath opening (WxL)
cm
23x14
24x30
Bath depth
cm
20
15
Filling volume from ... to
liters
12 ... 16
14 ... 20
Weight kg
43
41
Ambient temperature
°C
5 ... 40
5 ... 40
Mains power connection 230 V/50 Hz
V/ Hz
207-253 / 50
207-253 / 50
Current input (at 230 V)
A
12
12
Mains power connection 230 V/60 Hz
V/ Hz
207-253 / 60
207-253 / 60
Current input (at 230 V)
A
12
13
Mains power connection 115 V/60 Hz
V/ Hz
103-127 / 60
103-127 / 60
Current input (at 115 V)
A
15
14
Mains power connection 100 V/50-60 Hz V/
Hz 90-110V/50-60Hz ------
Current input (at 100 V)
A
15
------
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.