JULABO F34-EH Refrigerated and Heating Circulators User Manual
Page 22
Preparations
22
This may result in:
Burning of skin
Difficulties in breathing due to hot atmosphere
Safety recommendations
Employ suitable connecting tubing.
Make sure that the tubing is securely attached.
Avoid sharp bends in the tubing, and maintain a sufficient distance from
surrounding walls.
Regularly check the tubing for material defects (e.g. for cracks).
Preventive maintenance: Replace the tubing from time to time.
5.5.
Adjusting the pump flow
1
2
The pump flow is pre-adjusted in the factory and can be
modified to suit user requirements.
Using a screwdriver turn the screw (1) anti-clockwise by 360
°.
Using flat pliers turn the marking of the slide (2) to the
desired position.
Tighten the screw.
Examples:
extern 80 %
intern 100 %
intern 20 %
A
B
D
C
Internal applications in the bath
A 100 % internal bath circulation (for large bath tanks)
B Reduced internal bath circulation
(for smooth surface of bath fluid)
External/internal applications
C 40 % external discharge,
60 % internal circulation
(for large bath tanks)
D 80 % external discharge,
20 % internal circulation
(for small bath tanks)