JULABO F26-ED Refrigerated / Heating Circulator User Manual
Page 20
Preparations
20
Warning:
Tubing:
At high working temperatures the tubing used for temperature application and
cooling water supply represents a danger source.
A damaged tubing line may cause hot bath fluid to be pumped out within a
short time.
This may result in:
Burning of skin
Difficulties in breathing due to hot atmosphere
Safety recommendations
Employ suitable connecting tubing.
Make sure that the tubing is securely attached.
Avoid sharp bends in the tubing, and maintain a sufficient distance from
surrounding walls.
Regularly check the tubing for material defects (e.g. for cracks).
Preventive maintenance: Replace the tubing from time to time.
5.6.
Adjusting the pump flow
1
2
The pump flow is pre-adjusted in the factory and can be
modified to suit user requirements.
Using a screwdriver turn the screw (1) anti-clockwise by
360 °.
Using flat pliers turn the marking of the slide (2) to the
desired position.
Tighten the screw.
Examples:
Internal applications in the bath
A 100 % internal bath circulation (for large bath tanks)
B Reduced internal bath circulation
(for smooth surface of bath fluid)
External/internal applications
C 40 % external discharge,
60 % internal circulation (for large bath tanks)
D 80 % external discharge,
20 % internal circulation (for small bath tanks)