Safety notes for the user
12
Do not stay in the area below the unit.
Make sure you read and understand all instructions and safety precautions
listed in this manual before installing or operating your unit.
Never operate the unit without bath fluid in the bath.
Check the filling level of the bath fluid from time to time. Pump and heater
must always be fully covered with the bath fluid!
The instrument is not suited for unsupervised continuous operation.
Do not drain the bath fluid while it is hot!
Check the temperature of the bath fluid prior to draining (by switching the
unit on for a short moment for example).
Observe the limited working temperature range when using plastic bath
tanks.
Employ suitable connecting tubing.
Make sure that the tubes are securely attached.
Never operate damaged or leaking equipment.
Always turn off the unit and disconnect the mains cable from the power
source before performing any service or maintenance procedures, or
before moving the unit.
Always empty the bath before moving the unit.
Never operate equipment with damaged mains power cables.
Some parts of the bath cover and the pump connections may become
extremely warm during continuous operation. Therefore, exercise particular
caution when touching these parts.
Notice:
Check the safety installations at least twice a year!
Excess
temperature
protection
according to IEC 61010-2-010.
With a screwdriver turn back the adjustable excess temperature protection
until the shut-down point (actual temperature).
Low level protection according to IEC 61010-2-010.
To check the function of the float, it can be manually lowered with a