Preparations
18
5.3.
Filling / Draining
Notice:
Filling
Check the filling level of the bath fluid from time to time.
Pump and heater
must always be fully covered with the bath fluid!
The instrument is not suited for unsupervised continuous operation.
Filling
Take care that no liquid enters the interior of the circulator.
Recommended maximum filling level with water as bath liquid:
20 mm below the tank rim
After filling, immerse the samples in the bath or place the lid on
the bath, in case the opening is not to be used.
12
11
Draining:
Turn off the circulator and cooling machine.
Hold the venting grid, pull out and remove.
Slide a short piece of tube onto the drain port
(11) and hold it into a pail.
(only F25, F26)
Unscrew the drain tap (12) and empty the unit
completely.
Tighten the drain tap.
Notice:
Draining
Do not drain the bath fluid while it is hot!
Recommendation: Temperature range 5 °C to 40 °C
Check the temperature of the bath fluid prior to draining (by switching the unit
on for a short moment, for example).