JULABO ED-5 Open Bath Circulators User Manual

Page 15

background image

ED

15

See website for list of recommended bath fluids.
Contact: see page 5

Notice:

Please contact JULABO before using other than recommended bath fluids.
Only use non-acid and non corroding bath fluids.

JULABO takes no responsibility for damages caused by the selection of an
unsuitable bath fluid.
Unsuitable bath fluids are liquids which e.g.
 are very highly viscous

(much higher than 30 mm

2

/s at the respective working temperature)

 have corrosive characteristics or

 tend to cracking.


5.3.

Filling / Draining

Notice:

Filling

Check the filling level of the bath fluid from time to time.

Pump and heater

must always be fully covered with the bath fluid!

The instrument is not suited for unsupervised continuous operation.

Filling
 Take care that no liquid enters the interior of the circulator.
 Recommended maximum filling level with water as bath liquid:

20 mm below the tank rim

After filling, immerse the samples in the bath or place the lid on
the bath, in case the opening is not to be used.


Draining

 Press the mains switch to turn the circulator off.
 Remove the circulator from the bath tank.
 Empty the bath tank.

Notice:

Draining

Do not drain the bath fluid while it is hot!
Recommendation: Temperature range 5 °C to 40 °C

Check the temperature of the bath fluid prior to draining (by switching the unit
on for a short moment, for example).