beautypg.com

Bacharach RDM800 User Manual

Page 14

background image


HGM300 Refrigerant Gas Leak Monitor

NDIR Bench Measurement Description







Pyroelectric Detector

Sample Flow

NDIR Source Dual Optical Bandwidth Filters




Gas passing through the bench absorbs some I-R energy. Each of the 24 gasses in the onboard library

has a very specific absorption signature defining the amount of energy any of these gasses will absorb in

relation to the concentration present at the time of measurement. In order to increase accuracy and

eliminate false alarms a narrow (8.1 to 9.4 uM) optical bandwidth filter is used to limit the optical
spectrum to only those gasses listed in our library. All gasses outside of this bandwidth are ignored. The

output voltage from the pyroelectric detector is processed to eliminate any extraneous noise, amplified,

and digitized. This information is then sent to the onboard microcontroller, which analyzes and displays

the data. The unit is purged automatically which brings in a clean air sample. This air sample contains no

refrigerant gas, therefore, does not absorb any I-R energy. The energy measured in this purge sample is
used as a reference or baseline, when calculating PPM values for actual input channels, which allows us

to compensate for any changes that may occur over time in the measurement electronic circuitry, I-R

source, or I-R detector. This also eliminates for field calibration of the unit. All measurements are

temperature and pressure compensated to standard conditions to eliminate reading variability due to the

climate or altitude of the unit.


What specifically makes the HGM300 Single-Beam-Single-Wavelength NDIR measuring bench so

different?

1.) We gold plate the inside of the sampling tube. Gold is virtually untarnishable so this improves the

reflecting properties of the optical path, which improves overall measurement bench sensitivity and
accuracy.

2.) Because the optical properties have been improved we can reduce lamp intensity thereby improving

the I-R source life, which improves long-term reliability. We estimate source life to be beyond 5

years.

3.) I-R source modulation is required of all such devices. We do this electronically thereby eliminating

any mechanical parts and improving long-term reliability.

4.) Sample temperature changes can directly affect short-term measurement stability. Our auto

rezeroing function engages every 10 minutes or whenever the sample temperature changes by more

than 0.2 deg. C, or the enclosure interior temperature changes by more than 2.0 deg. C. Thereby,

improving measurement stability and accuracy, and completely eliminating the effects of sample-

component temperature drift anywhere in the monitor.

Rev 3/2/09

14