beautypg.com

Dakota Ultrasonics PR-82 User Manual

Page 8

background image

Dakota Ultrasonics

4

The PRB-0 key is used to "zero" the PR-8

2

in much the same way that

a mechanical micrometer is zeroed. If the tool is not zeroed correctly, all of

the measurements that the tool makes may be in error by some fixed value.

Refer to page 12 for an explanation of this important procedure.

The CAL key is used to enter and exit the PR-8

2

's calibration mode.

This mode is used to adjust the sound-velocity value that the PR-8

2

will use

when calculating thickness. The tool will either calculate the sound-velocity

from a sample of the material being measured, or allow a known velocity

value to be entered directly. Refer to page 13 for an explanation of the two

CAL functions available.

The IN/MM key is used to switch back and forth between English and

metric units. This key may be used at any time, whether the tool is

displaying a thickness (IN or MM) or a velocity value (IN/

µs or M/s).