PX-7 Modes

Scan Mode

While the PX-7 excels at making single point measurements, it is

sometimes desirable to examine a larger region, searching for the thinnest

point. The PX-7 includes a feature, called Scan Mode, which allows it to do

just that.

In normal operation, the PX-7 performs and displays four

measurements every second, which is quite adequate for single

measurements. In Scan Mode, however, the tool performs eight

measurements every second, but does not display them. While the

transducer is in contact with the material being measured, the PX-7 is

keeping track of the lowest measurement it finds. The transducer may be

"scrubbed" across a surface, and any brief interruptions in the signal will be

ignored. When the transducer loses contact with the surface for more than

a second, the PX-7 will display the smallest measurement it found.

When the PX-7 is not in calibration mode, press the SCAN key to turn

Scan Mode on and off. A brief message will appear in the display

confirming the operation. While scanning, the display will show a moving

series of dashes instead of a thickness value. When the transducer is

removed from the material being scanned, the PX-7 will (after a brief

pause) display the smallest measurement it found.