Page 25: Sensor head inspections -2, 3 cables, Inspections -2, Problems found during inspection -2

20909-01C Rev 04
Page 10-2
10.2.2
Sensor Head Inspections
Verify the absence of the following:
Loose or damaged latch hardware
Loose or damaged boot seal clamps
Damage to enclosure
Damage to seals
Loose or damaged access cover (including screws and seals)
Loose
handles
Loose or damaged electrical connector or seals at the
connector
Evidence from the condition of the pipe that a pipe leak under
the Sensor Head might occur before the next inspection
(reference paragraph 7.6).
10.2.3
Cables Inspections
Verify the absence of the following:
Damaged
cables
Improperly
installed
cables (for example, cable is under tension
or too tight bend radius)
Loose cable management elements (strain reliefs, clamps, etc.)
10.3
Problems Found During Inspection
If inspections turn up serious problems, remove the sonar process
monitoring system from service until repairs can be made.
.