EBU-222 Instruction Manual, version 7/28/2011

5

www.cbsscientific.com

SECTION 1

General Information

1.1 Introduction

Electroblotting is a faster, easier, and more efficient method to permanently transfer PAGE gels to nitrocellulose

and nucleic acids to nylon, than capillary (passive diffusion) blotting. For a similar transfer, traditional “capillary

blotting” can take 6-24 hours to complete compared to only 2 hours or less using electroblotting. The C.B.S. Sci-

entific Electrophoretic Blotting Systems come equipped with; two electrode panels, one color-coded gel cassette

assembly, electrophoresis tank with cooling chamber and safety cover with attached power leads. The EBU-222

gel cassette size is 20cm x 21.5cm and can transfer up to three gels at one time. The EBU-222 gel cassette size

is 9 x 11cm and can transfer up to 2 gels at one time.

1.2 Specifications

Constructions:

Buffer chamber, safety cover

Acrylic

Electrode panels

Acrylic and platinum wire .010” diameter

Power cords

FR Polypropylene/Silicone, rated 7500V, 200mA, 65° C

Gel cassette

Perforated acrylic panels red and black, two

sponge

pads

Safety Certification

EN61010-1-1993 (IEC1010-1)

Shipping

Weight 5

lbs.

Overall Size (l)x(w)x(h) cm

19 x 11 x 28.5

Maximum Gel Size - cm

18 x18

Distance between electrodes -cm

8.7

Recommended buffer volume

1 cassette

3.5

liters

2 cassettes

3.0

liters

1.3 Safety

Power to the Electro-Blotter is to be supplied by an external DC voltage power supply that must be ground

isolated so that the DC voltage output floats with respect to ground. For any power supply used, the maximum

specified operating parameters for the units are:

Maximum Limits

100

VDC

70

watts

power

1500mA current

50ºC ambient temperature

Current to the unit, provided from the external power supply, must enter the unit through the lid assembly, provid-

ing a safety interlock to the user. Current to the unit is broken when the lid is removed. Do not attempt to use the

unit without the safety lid, and always turn the power supply off before removing the lid, or when working with the

unit in any way. Follow safety precautions specified by the power supply manufacturer.