Exposure Limit Value (ELV)
The table indicates that the operator using 15 ml Falcon tubes in touch vortex mode (3500 rpm) with reported
usage would reach the ELV in 444 minutes and using 50 ml Falcon tubes in 136 minutes. Based on a usage
pattern of 30 seconds vortex / 5 seconds gap, this means that approximately 761 of the 15 ml or 233 of the 50
ml Falcon tubes could produce exposures that reach the ELV. Immediate action must thus be taken to limit
exposures to below the ELV. With a usage pattern of 15 seconds vortex / 5 seconds gap, 1332 x 15 ml tubes
or 408 x 50 ml tubes must be vortexed until the ELV is reached. With a usage pattern of 45 seconds vortex / 5
seconds gap 532 x 15 ml tubes or 163 x 50 ml tubes must be vortexed.
Number of tubes to reach
Vortex time
(5 sec gap)
EAV ELV
15 ml
EAV ELV
50 ml
15 sec
333
1332
102
408
30 sec
190
761
58
233
45 sec
133
532
40
163
Number of tubes for not exceeding
Vortex
time
(5 sec gap)
EAV ELV
15 ml
EAV ELV
50 ml
15 sec
333
1332
102
408
30 sec
190
761
58
233
45 sec
133
532
40
163