7 Maintenance

7.1 Cleaning

7.1.1

Cleaning the device and accessories

Clean the housing of the MixMate, the vortex mat, the plate holder and the tube holders regularly.

Too fast

Mixing load not properly positioned

in the plate holder.

Check that the mixing load is

fixed in position.

Too fast

Continuous vortexing at

frequencies > 2000 rpm.

Reduce the continuous

vortexing frequency to

≤ 2000 rpm.

ERR00 - ERR03 /

ERR06 - ERR11

Electronics error

Switch off the device and

switch it on again after 5 s.

ERR04 - ERR05

Device overheats.
The ventilation slots on the

underside of the device are

blocked.

1. Switch off the device and leave

to cool down for 10 minutes.

2. Ensure that the ventilation slots

on the underside of the device

are clear.

3. Switch the device back on

again.

ERR12

Incorrect power supply

Make sure that the supply

voltage and the power

frequency are compatible with

the information given on the

device name plate. This is

located on the underside of the

device.

ERR13 Software

error

Switch off the device and

switch it on again after 5 s.

ERR15 and ERR16 Hardware

error

Contact your Eppendorf

partner.

Symptom/

message

Cause

Remedy

DANGER!

DANGER! Electric shock as a result of penetration of liquid.

Switch off the device and disconnect it from the power supply before starting

cleaning or disinfecting.

Do not allow any liquids to penetrate the inside of the housing.

Do not spray clean/spray disinfect the housing.

Wait until the device is completely dry before connecting it to the power source

again.

EN